Voltage Test - Fluke 8062A Instruction Manual

True-rms multimeter
Table of Contents

Advertisement

4-14. Voltage Test

Use the following procedure to verify proper operation of the dc and ac
voltage functions.
1.
Connect the UUT and the DMM Calibrator as shown in Figure 4-4.
2.
For each step in Table 4-2, select the UUT function and range as
indicated. Program the DMM Calibrator for the specified input signal
and verify that the displayed UUT value is within the indicated limits.
Step
Function
1
DC Voltage
2
3
4
5
6
7
AC Voltage
8
9
AC Voltage
10
11
12
13
14
15
16
AC Voltage
17
18
19
AC Voltage
20
21
22
23
Table 4-2. Voltage Test
Range
Level
200 mV
+ 190.00 mV
200 mV
-190.00 mV
2V
1.9000V
20V
19.000V
200V
190.00V
1000V
1000V
200 mV
100.00 mV rms
2V
1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
200.0 mV rms
200.0 mV rms
20V
10.000V rms
10.000V rms
10.000V rms
200V
100.00V rms
200V
100.00V rms
200V
100.00V rms
750V
750.0V rms
750V
750.0V rms
Performance Tests
Input Signal
Freq.
dc
200 Hz
20 kHz
200 Hz
1 kHz
10 kHz
30 kHz
20 kHz
200 Hz
30 kHz
200 Hz
10 kHz
30 kHz
200 Hz
10 kHz
30 kHz
100 Hz
1 kHz
Maintenance
Display
Limits
+189.90 to 190.11
-189.89 to -190.11
1.8989 to 1.9011
18.985 to 19.015
189.85 to 190.15
999.1 to 1000.9
99.40 to 100.60
98.60 to 101.40
.9940 to 1.0060
.9930 to 1.0070
.9930 to 1.0070
.9860 to 1.0140
.9890 to 1.0110
.1978 to .2022
.7940 to .2060
9.940 to 10.060
9.480 to 10.520
9.460 to 10.540
99.40 to 100.60
94.80 to 105.20
94.60 to 105.40
734.0 to 766.0
734.0 to 766.0
4-13