Fluke 8062A Instruction Manual page 17

True-rms multimeter
Table of Contents

Advertisement

Electrical
The following specifications are based on a one-year calibration cycle, an
operating temperature of 18 to 28°C (64 to 82°F) and a relative humidity not
exceeding 90%.
DC Voltage
Range
200 mV
2V
20V
200V
1000V
Response Time.............................. 1 second maximum, to rated
Input Impedance ............................ 10 M shunted by >100 pf
Normal Mode Noise Rejection ....... >60 dB at 50 Hz or 60 Hz
Common Mode Noise Rejection .... >120 dB at dc, >90 dB at 50 Hz and
Overload Protection ....................... 1000V dc or peak ac continuous,
DC Voltage, High Impedance Mode
All specifications are the same as for the dc voltage mode except the
following (only 200 mV and 2V ranges are available):
Range
200 mV
2V
Input Impedance ............................ >1000 M , typically 10,000 M
Overload Protection ....................... 300V dc or rms continuous, 20
Introduction and Specifications
Table 1-2. 8062A Specifications
Resolution
.01 mV
.1 mV
1 mV
10 mV
100 mV
Resolutions
.01 mV
.1 mV
Accuracy
±(% of reading + no. of digits)
0.05% + 2
0.07% + 2
accuracy within selected range.
60 Hz (1 k imbalance)
except 20 seconds maximum on 200
mV and 2V ranges above 300V dc or
rms.
Accuracy
±(% of reading + no. of digits)
0.06% + 2
seconds maximum 300V to 1000V dc
or peak ac.
Specifications
1
1-5