Equipment Connection For Current Test - Fluke 8062A Instruction Manual

True-rms multimeter
Table of Contents

Advertisement

8062A
Instruction Manual
2.
For each step in Table 4-4, select the UUT function and range as
indicated. Program the DMM Calibrator for the specified input signal
and verify that the displayed UUT value is within the indicated limits.
Figure 4-5. Equipment Connection for Current Test
Step
Function
1
2
3
4
Current dc
5
6
7
8
Current ac
4-16
UUT
Table 4-4. Current Test
Range
Level
200 A
190.00 A
200 A
-190.00 A
2 mA
1.9000 mA
20 mA
19.000 mA
200 mA
190.00 mA
2000 mA
1900.0 mA
2000 mA
-1900.0 mA
20 mA rms
19.000 mA
Input Signal
Freq.
dc
dc
dc
dc
dc
dc
dc
1 kHz
DMM
Calibrator
HI
LO
dy36f.eps
Display
Limits
189.41 to 190.59
-189.41 to -190.59
1.8941 to 1.9059
18.941 to 19.059
188.65 to 191.35
1886.5 to 1913.5
-1886.5 to -1913.5
18.848 to 19.152