Disassembling The Lcd - Fluke 8062A Instruction Manual

True-rms multimeter
Table of Contents

Advertisement

8062A
Instruction Manual
The LCD interconnect (item 7) and the microcomputer interconnect
(item 5) should not be touched with fingers or contaminated. Handle
these items with tweezers and keep them clean.
The microcomputer interconnect (item 7) is susceptible to corrosion
caused by the reaction between the metal in the connector and possible
contaminates in the air such as smoke or sulfur. Store the connector in
an air-tight container if the LCD is disassembled for a long period of
time.
Do not get fingerprints or dirt on the LCD display, the display lens, or
the gasket.
While the LCD and microcomputer pcb are assembled, take care not to
press down on the display lens because pressure could damage the LCD.
To disassemble the LCD, use your thumbnails and push on the corners of the
LCD display, gasket, and display lens so that all three components slide out
together as shown in Figure 4-3.
It is not necessary to remove the main pcb from the bottom case to
disassemble or reassemble the LCD.
To assemble the LCD, use the following procedure:
1.
Align the LCD display (item 8) as indicated in Figure 4-2 and slide it
into place. The bottom edge of the LCD display should compress the
LCD interconnect (item 7) and slide underneath the two plastic notches
on the LCD bracket (item 1).
2.
Refer to Figure 4-2 and follow steps 9 and 10 to complete assembly.
4-10
Slide Out
Figure 4-3. Disassembling the LCD
Note
Push corners with thumbnails.
dy34c.eps