Main Pcb Access; Calibration And Backup Fuse (F2) Access - Fluke 8062A Instruction Manual

True-rms multimeter
Table of Contents

Advertisement

Green Power
Switch Cap
Pry fuse out from the side.
Figure 4-1. CALIBRATION AND BACKUP FUSE (F2) ACCESS
4-7.

Main PCB Access

Use the following procedure to gain access to the main pcb:
1.
Remove the screw in the center of the ac shield and remove the shield.
2.
Using your index finger, lift up the lower right corner of the main pcb
until it is free. Then pull the pcb to the right until it clears the shelf under
the buttons.
Elastomeric Strip
Maintenance
General Information
AC Shield
Remove before removing
Main PCB. When reassembling,
install shield after installing
Main PCB.
Backup Fuse F2
4
dy32c.eps
4-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents