Advertisement
The Fluke Model 110, Model 113, Model 114, Model 115, and Model 117 (the Meter or Product) are batterypowered, true-rms multimeters with a 6000-count display and a bar graph. This manual applies to all models. All figures show the Model 117 unless indicated.
For Product safety information, see the printed 110/113/114/115/117
Safety Information included with the Product or located on the Fluke website.
To alert you to the presence of a potentially hazardous voltage, the Z symbol is displayed when the Meter measures a voltage ≥30 V or a voltage overload (OL) condition. When making frequency measurements >1 kHz, the Z symbol is unspecified.
Personal injury or damage to the Meter can occur if you attempt to make a measurement with a lead in an incorrect terminal.
To remind you to check that the test leads are in the correct terminals, LEAd is briefly displayed and an audible beep sounds when you move the rotary switch to or from any A (Amps) position.
The manual explains features for multiple models. Because models have different features, not all of the information in the manual may apply to your Meter. Use Table 1 to identify the features of your Meter.
Table 1 is a list of the features for each Meter.
Table 1. Features
Switch Position | Measurement Function | 110 | 113 | 114 | 115 | 117 |
OFF | The Meter is turned off. | |||||
AUTO-V LoZ | Automatically selects ac or dc volts based on the sensed input with a low impedance input. | |||||
AC voltage from 0.06 V to 600 V. Frequency from 5 Hz to 100 kHz. | ||||||
DC voltage from 0.001 V to 600 V. | ||||||
AC voltage from 6.0 to 600 mV, dc-coupled. DC voltage from 0.1 to 600 mV. | ||||||
Ω | Ohms from 0.1 Ω to 40 MΩ. | |||||
Continuity beeper turns on at <20 Ω and turns off at >250 Ω. | ||||||
LoZ low impedance measurement function to simultaneously test for voltage or continuity. | ||||||
Diode Test. Displays OL above 2.0 V. | ||||||
Farads from 1 nF to 9999 μF. | ||||||
AC current from 0.1 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off). >10.00 A display flashes. >20 A, OL is displayed. DC-coupled. Frequency from 45 Hz to 5 kHz. | ||||||
DC current from 0.001 A to 10 A (>10 A to 20 A, 30 seconds on, 10 minutes off). >10.00 A display flashes. >20 A, OL is displayed. | ||||||
Volt Alert | Non-contact sensing of ac voltage. | |||||
Note: All ac functions and Auto-V LoZ are true-rms. AC voltage is ac-coupled. Auto-V LoZ, AC mV and AC amps are dc-coupled. |
Table 2 is a list of the features for each display.
Table 2. Display
No. | Symbol | Meaning | Model |
1 | Volt Alert | The Meter is in the VoltAlert™ non-contact voltage detect mode. | 117 |
2 | The Meter function is set to Continuity. | 110, 113, 114, 115, 117 | |
3 | The Meter function is set to Diode Test | 113, 115, 117 | |
4 | Input is a negative value. | 110, 113, 114, 115, 117 | |
5 | X Unsafe voltage. Measured input voltage ≥30 V, or voltage overload condition (OL). | 110, 113, 114, 115, 117 | |
6 | Display hold enabled. Display freezes present reading. | 110, 113, 114, 115, 117 | |
7 | MIN MAX AVG mode enabled. Maximum, minimum, average or present reading displays | 110, 113, 114, 115, 117 | |
8 | (Red LED) | Voltage presence through the non-contact VoltAlert sensor | 117 |
9 | LoZ | The Meter is measuring voltage or capacitance with a low input impedance. | 113,114, 115, 117 |
10 | nμF mV μA MkΩ kHz | Measurement units. | 110, 114, 115, 117 |
11 | DC AC | Direct current or alternating current | 110, 113, 114, 115, 117 |
12 | Battery low warning. | 110, 113, 114, 115, 117 | |
13 | 610000 mV | Indicates the Meter's range selection. | 110, 114, 115, 117 |
14 | (Bar graph) | Analog display. | 110, 113, 114, 115, 117 |
15 | Auto Volts | The Meter is in the Auto Volts function. | 114, 117 |
Auto | Autoranging. The Meter selects the range for best resolution. | 110, 113, 114, 115, 117 | |
Manual | Manual ranging. User sets the Meter's range. | 110, 113, 114, 115, 117 | |
16 | + | Bar graph polarity | 110, 113, 114, 115, 117 |
17 | The input is too large for the selected range. | 110, 113, 114, 115, 117 | |
18 | Test lead alert. Briefly displayed whenever the Meter's function switch is rotated to or from any A position. | 115, 117 |
Table 3 is a list of terminals on the Meter.
Table 3. Terminals
No. | Description | Model |
1 | Input terminal for measuring ac and dc current to 10 A. | 115, 117 |
2 | Common (return) terminal for all measurements. | 110, 113, 114, 115, 117 |
3 | Input terminal for measuring voltage, continuity, resistance, capacitance, frequency and testing diodes. | 110, 113, 114, 115, 117 |
Table 4 is a list of error messages for the Meter.
Table 4. Error Messages
Error Messages | |
Battery must be replaced before the Meter will operate. | |
Calibration required. Meter calibration is required before the Meter will operate. | |
Internal error. The Meter must be repaired before it will operate. | |
Internal error. The Meter must be repaired before it will operate. |
If the Meter is ON, but inactive and not connected to voltage for more than 20 minutes, the display goes blank to save battery life. To use the Meter, press any button or turn the rotary switch. To disable the Sleep mode, see Power-Up Options. The Sleep mode is always disabled in the MIN MAX AVG mode.
The MIN MAX AVG recording mode captures the minimum and maximum input values (ignoring overloads), and calculates a running average of all readings. When the Meter detects a new high or low, the Meter beeps.
Note
Autoranging and Battery Saver™ are disabled in MIN MAX AVG mode.
To avoid electric shock, when Display HOLD is activated, be aware that the display will not change when you apply a different voltage.
In the Display HOLD mode, the Meter freezes the display.
Push to toggle the backlight on and off.
The backlight automatically turns off after 40 seconds. To disable backlight auto-off, see Power-Up Options.
The Meter has both Manual and Autorange modes. The Meter defaults to Autorange. To toggle between Manual and Autorange, push for 1 second.
Note
You cannot manually change the range in the MIN MAX AVG or Display HOLD modes. If you push while in MIN MAX AVG or Display Hold, the Meter beeps twice, indicating an invalid operation and the range does not change.
To select a Power-Up Option, hold down the button indicated in Table 5 while turning the Meter from OFF to any other function. Power-Up Options are canceled when you turn off the Meter and when sleep mode is activated.
Table 5. Power-Up Options
Button | Power-Up Options |
Turns on all display segments until button is released. | |
Disables beeper. shows when enabled. | |
113 - Turns on all display segments until button is released. | |
115, 117 - Enables low impedance capacitance measurements. shows when enabled. | |
Disables Battery Saver™ (Sleep mode). shows when enabled. | |
Disables auto backlight off. is displayed when enabled. |
When connecting the test leads to the circuit or device, connect the common (COM) test lead before connecting the live lead; when removing the test leads, remove the live lead before removing the common test lead.
To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
Note
Use the continuity function as a fast, convenient method to check for opens and shorts. For maximum accuracy in making resistance measurements, use the Meter's resistance () function.
With the function switch in the position, the Meter automatically selects a dc or ac voltage measurement based on the input applied between the V or + and COM jacks.
This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to ghost voltages.
With the function switch in the position, the Meter measures ac plus dc millivolts. Press to set the Meter to dc millivolts.
To avoid personal injury or damage to the Meter:
To measure current:
The millivolt and voltage function of the Meter can be used with an optional mV/A output Current Probe to measure currents that exceed the rating of the Meter. Make sure the Meter has the correct function selected, AC or DC, for your current probe. Refer to a Fluke catalog or contact your local Fluke representative for compatible current clamps.
To avoid electrical shock, disregard the bar graph for frequencies >1 kHz. If the frequency of the measured signal is >1 kHz, the bar graph and are unspecified.
The Meter measures the frequency of a signal by counting the number of times the signal crosses a trigger level each second. The trigger level is 0 V, 0 A for all ranges.
Press to turn on or turn off the frequency measurement function on and off. Frequency works with ac functions only.
In frequency, the bar graph and range annunciator indicate the ac voltage or current present.
Select progressively lower ranges using manual ranging for a stable reading.
To detect the presence of ac voltage, place the top of the Meter close to a conductor. The Meter gives an audible as well as visual indication when voltage is detected. The sensitivity settings are:
The VoltAlert detector works in bare wire applications with voltages as low as 24 V in the Hi setting.
If there is no indication, voltage could still be present. Do not rely on the VoltAlert detector with shielded wire. Operation may be effected by differences in socket design, insulation thickness and type.
For making capacitance measurements on cables with ghost voltage:
In this mode, capacitance measurements will have a lower accuracy and lower dynamic range.
Note
This setting is not saved when the Meter is turned off or goes into sleep mode.
The bar graph is like the needle on an analog meter. It has an overload indicator ( ) to the right and a polarity indicator () to the left.
Because the bar graph is much faster than the digital display, the bar graph is useful for making peak and null adjustments.
The bar graph is disabled when measuring capacitance. In frequency, the bar graph and range annunciator indicates the underlying voltage or current up to 1 kHz.
The number of segments indicates the measured value and is relative to the full-scale value of the selected range.
In the 60 V range, for example (see below), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. An input of -30 V turns on the negative sign and the segments up to the middle of the scale.
Maintenance of the Meter consists of battery and fuse replacement, as well as case cleaning.
Test fuse as shown in Figure 1.
To avoid shock, injury, or damage to the Meter:
See Figure 2 for disassembly.
To remove the battery door for battery replacement:
To open the case for fuse replacement:
Wipe the case with a damp cloth and mild detergent. Dirt or moisture in the terminals can affect readings.
Accuracy is specified for 1 year after calibration, at operating temperatures of 18°C to 28°C, with relative humidity at 0% to 90%.
Extended specifications are available at www.Fluke.com.
Maximum voltage between any terminal and earth ground | 600 V |
Fuse for A input (115 & 117 only) | 11 A, 1000 V, IR 17 kA |
Display | |
Digital | 6000 counts, updates 4/s |
Bar Graph | 33 segments, updates 32/s |
Temperature | |
Operating | -10°C to 50°C |
Storage | - 40°C to 60°C |
Temperature Coefficient | 0.1 x (specified accuracy)/°C (<18°C or >28°C) |
Altitude | |
Operating | 2000 meters |
Storage | 10 000 meters |
Relative Humidity | 95% to 30°C, 75% to 40°C, 45% to 50°C |
Battery | IEC 6LR61 |
Battery Life | |
113 | Alkaline: 300 hours typical, without backlight |
110, 114, 115, 117 | Alkaline: 400 hours typical, without backlight |
Safety | IEC 61010-1: Pollution Degree 2 IEC 61010-2-033 |
113 | Measurement CAT IV 600 V |
110, 114 | Measurement CAT III 600 V |
115, 117 | Measurement CAT III 600 V, 10 A |
Ingress Protection | IEC 60529: IP42 (non-operating) |
Electromagnetic Compatibility (EMC) | |
International | IEC 61326-1: Portable Electromagnetic Environment CISPR 11: Group 1, Class A |
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Korea (KCC)
Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it.
This equipment is intended for use in business environments and not to be used in homes.
USA (FCC)
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Table 6. Accuracy Specifications
Function | Range | Resolution | Accuracy ± ([% of Reading] + [Counts]) | Model | |||||
DC Millivolts | 600.0 mV | 0.1 mV | 0.5% + 2 | 110, 114, 115, 117 | |||||
DC Volts | 6.000 V | 0.001 V | 0.5% + 2 | 110, 114, 115, 117 | |||||
60.00 V | 0.01 V | ||||||||
600.0 V | 0.1 V | ||||||||
DC, 45 to 500 Hz | 500 Hz to 1 kHz | ||||||||
Auto-V LoZ[1] True-rms | 600.0 V | 0.1 V | 2.0% + 3 | 4.0% + 3 | 114, 117 | ||||
CHEK[4] | 6.000 V | 0.001 V | 2.0% + 3 | 4.0% + 3 | 113 | ||||
60.00 V | 0.01 V | ||||||||
600.0 V | 0.1 V | ||||||||
45 to 500 Hz | 500 Hz to 1 kHz | ||||||||
AC millivolts[1] True-rms | 600.0 mV | 0.1 mV | 1.0% + 3 | 2.0% + 3 | 110, 114, 115, 117 | ||||
AC Volts[1] True-rms | 6.000 V | 0.001 V | 1.0% + 3 | 2.0% + 3 | 110, 114, 115, 117 | ||||
60.00 V | 0.01 V | ||||||||
600.0 V | 0.1 V | ||||||||
Continuity[5] | 600 Ω | 1 Ω | Beeper on <20 Ω, off >250 Ω. Detects opens or shorts of 500 μs or longer. | 110, 114, 115, 117 | |||||
--- | --- | 113 | |||||||
Ohms[5] | 600.0 Ω | 0.1 Ω | 0.9% + 2 | 110, 113, 114, 115,117 | |||||
6.000 kΩ | 0.001 kΩ | 0.9% + 1 | |||||||
60.00 kΩ | 0.01 kΩ | 0.9% + 1 | |||||||
600.0 kΩ | 0.1 kΩ | 0.9% + 1 | 110, 114, 115, 117 | ||||||
6.000 MΩ | 0.001 MΩ | 0.9% + 1 | |||||||
40.00 MΩ | 0.01 MΩ | 5.0% + 2 | |||||||
Diode Test[5] | 2.000 V | 0.001 V | 0.9% + 2 | 115, 117 | |||||
2.0% + 3 | 113 | ||||||||
Capacitance[5] | 1000 nF | 1 nF | 1.9% + 2 | 113, 115, 117 | |||||
10.00 μF | 0.01 μF | 1.9% + 2 | |||||||
100.0 μF | 0.1 μF | 1.9% + 2 | |||||||
9999 μF | 1 μF | 100 μF - 1000 μF: 1.9% +2 >1000 μF: 5% + 20 | |||||||
Lo-Z Capacitance (Power-up option) | 1 nF to 500 μF | 10% + 2 typical | 115, 117 | ||||||
AC Amps True-rms[1] (45 Hz to 500 Hz) | 6.000 A | 0.001 A | 1.5% + 3 | 115, 117 | |||||
10.00 A[3] | 0.01 A | ||||||||
DC Amps | 6.000 A | 0.001 A | 1.0% + 3 | 115, 117 | |||||
10.00 A[3] | 0.01 A | ||||||||
Hz (V or A input)[2] | 99.99 Hz | 0.01 Hz | 0.1% + 2 | 115, 117 | |||||
999.9 Hz | 0.1 Hz | ||||||||
9.999 kHz | 0.001 kHz | ||||||||
50.00 kHz | 0.01 kHz | ||||||||
99.99 kHz | 0.01 kHz | ||||||||
Notes:
|
Table 7. Input Characteristics (110, 114, 115, 117)
Function | Input Impedance (Nominal) | Common Mode Rejection Ratio (1 kΩ Unbalanced) | Normal Mode Rejection | |
Volts AC | >5 MΩ <100 pF | >60 dB at dc, 50 or 60 Hz | 0.5% + 2 | - - - |
Volts DC | >10 MΩ <100 pF | >100 dB at dc, 50 or 60 Hz | 0.5% + 2 | - - - |
Auto-V LoZ | ~3 kΩ <500 pF | >60 dB at dc, 50 or 60 Hz | - - - | |
Open Circuit Test Voltage | Full Scale Voltage | Short Circuit Current | ||
Ohms | <2.7 V dc | to 6.0 MΩ | 40 MΩ | <350 μA |
<0.7 V dc | <0.9 V dc | |||
Diode Test | <2.7 V dc | 2.000 V dc | <1.2 mA |
Table 8. Input Characteristics (113)
Function | Input Impedance (Nominal) | Common Mode Rejection Ratio |
k CHEK | ~3 kΩ <300 pF | >60 dB at dc, 50 or 60 Hz |
Open Circuit Test Voltage | Full Scale Voltage | |
Ohms | <2.7 V dc | <0.7 V dc |
Diode Test | <2.7 V dc | <2.000 V dc |
Short Circuit Current | ||
Ohms | <350 μA | |
Diode Test | <1.0 mA |
MIN MAX Recording Accuracy and Response Time (113)
Specified accuracy of the measurement function ±40 counts in k CHEK for changes >500 ms in duration, ±12 counts in e for changes >325 ms in duration. Typical 100 ms response to 80%. Response time not specified for capacitance.
To contact Fluke, call one of the following telephone numbers:
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
To request a printed manual, visit www.fluke.com/productinfo.
© 2020 Fluke Corporation.
All rights reserved.
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Fluke 110/113/114/115/117 - True-rms Multimeter Manual
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the 110 that isn't answered in the manual? Leave your question here.