753/754
Users Manual
Square root percent of scale is in effect until the Product
is changed to a different source function or the Scale
softkey is pushed and you select a different scale mode.
Step and Ramp the Output Value
Two features are available that let you adjust the value of
source functions, except pressure. For pressure, an
external pressure source must be used:
•
Step the output manually with and , or in
automatic mode.
•
Ramp the output with optional continuity or V trip
detect.
Manual Step Use
The manual Step feature selects a step size in
engineering units (mV, V, mA, °C, etc.) or % of scale.
Step the output in % of scale to quickly jump between 0 %
and 100 % (set step size = 100 %) or 0-50-100 % (set
step size = 50 %). Step works in SOURCE and in
MEASURE/SOURCE modes.
To select a step size:
1.
See the applicable Source Mode subheading in this
manual (for example, "Source Electrical Parameters")
and connect the Product to the test circuit.
2.
If necessary, push for SOURCE mode.
3.
Set the Product for the necessary source value.
62
Shop for Fluke products online at:
4.
To step the source value in % of scale, set the % of
scale value as given before in "Measurement Scale".
5.
Push the Step softkey.
6.
Use the numeric keypad to record the step size in the
units shown on the display.
7.
Push the Done softkey.
8.
Push and to adjust the output in steps.
Auto Step Use
To configure the Product to make a sequence of steps
automatically, either once through the sequence or
repetitively:
1.
See the applicable Source Mode subheading in this
manual (for example, "Source Electrical Parameters")
and connect the Product to the test circuit.
2.
If necessary, push for SOURCE mode.
3.
Set the Product for the necessary source value.
4.
To step the source value in % of scale, set the % of
scale value as given before in "Measurement Scale".
5.
Push the Step softkey.
6.
Push the Auto Step softkey.