13.1 Information on electromagnetic compatibility (EMC)
Recommended minimum distance between portable and mobile RF communications devices
and the CuratOR LX490W
The CuratOR LX490W is intended for use in an electromagnetic environment in which interference
due to electromagnetic radiation is controlled.
The customer or user of the monitor can help prevent electromagnetic interference by maintaining
the recommended minimum distance between portable and mobile RF communications devices
(transmitter) and the monitor as shown below. This is based on the maximum output power of the
communication device.
Maximum nominal
Recommended minimum distance according to the frequency of the
output power of
transmitter (m)
the transmitter
150 kHz to 80 MHz
(W)
d = 0.583 √P
0.01
0.058
0.1
0.184
1
0.583
10
1.844
100
5.833
For transmitters whose maximum nominal output power is not shown above, the recommended mini-
mum distance "d" in meters (m) can be determined using the formula for calculating the frequency of
the transmitter. "P" here stands for the transmitter's maximum measured nominal output power in
watts (W), as recommended by the transmitter's manufacturer.
Note: For 80 MHz and 800 MHz, the recommended minimum distance for the higher frequency
range applies.
Note: This information may not be applicable in all situations. Absorption and reflection by struc-
tures, objects, and people impact the propagation of electromagnetic waves.
58
80 MHz to 800 MHz
d = 0.35 √P
0.035
0.110
0.35
1.106
3.5
800 MHz to 2.7 GHz
d = 0.7 √P
0.07
0.221
0.7
2.213
7
CuratOR LX490W