Information and manufacturer's declaration – electromagnetic radiation
The CuratOR LX490W is intended for use in the electromagnetic environments noted below.
The customer or user of the CuratOR LX490W has to ensure that the device is used in such an envi-
ronment.
Interference im-
IEC 60601
munity test
Measure-
ment level
Conducted distur-
3 Vrms
bances
150 kHz to
80 MHz
IEC/EN
61000-4-6
6 Vrms
ISM bands
High-frequency
3 V/m
electromagnetic
80 MHz to
fields
2.7 GHz
IEC/EN
61000-4-3
Note: The higher frequency range applies at 80 MHz and 800 MHz.
Note: This information may not be applicable in all situations. Absorption and reflection by struc-
tures, objects, and people impact the propagation of electromagnetic waves. The measurement lev-
els were established according to IEC60601-1-2 4th Edition - Professional Healthcare.
a
The field strengths of fixed transmitters, for example the base station for cordless and mobile tele-
phones, radio, land mobile radio, ham radio, and television cannot be determined precisely in ad-
vance. To evaluate the electromagnetic environment using fixed transmitters, an electromagnetic lo-
cation measurement should be included. If the measured field strength in the environment where the
device is used exceeds the applicable RF compliance level, observe the monitor to ensure its proper
operation. If improper operation is observed, in some circumstances additional measures may be
necessary, such as reorienting or repositioning the device.
b
The field strength beyond the frequency range 150 kHz to 80 MHz should be less than 3 V/m.
CuratOR LX490W
Instructions for Use, 06/2017
13.1 Information on electromagnetic compatibility (EMC)
Compliance
Information regarding the electromagnetic envi-
level
ronment
6 Vrms
Portable and mobile RF communications devices
may only be operated in the vicinity of the monitor
when in compliance with the recommended mini-
mum distance.
It is determined using the formula for calculating the
frequency of the transmitter.
Recommended minimum distance
10 V/m
d = [3.5/6] √P = 0.583 √P
d = [3.5/10] √P = 0.35 √P,
80 MHz to 800 MHz
d = [7/10] √P = 0.7 √P,
800 MHz to 2.7 GHz
In this case, "P" stands for the measured maximum
nominal output power in watts (W) of the transmitter
recommended by the transmitter manufacturer, and
"d" for the recommended minimum distance in me-
ters (m).
The field strengths of fixed transmitters according to
electromagnetic location measurement
less than the compliance level in each individual fre-
quency range
Interference can occur when used in the vicinity of
devices identified with the following symbol.
b
.
Appendix
a
have to be
57