Eizo CuratOR LX490W Instructions For Use Manual page 16

49” lcd monitor
Table of Contents

Advertisement

Setup and installation
4.1 Installation site
Avoid dusty environments
The monitor is intended for use in the clean environment of medical diagnostics. In dusty
environments, ventilation holes in the back can allow dust to penetrate into the monitor. In
the worst case, deposits are possible which become evident as dark spots in a white picture
and result in deterioration of the luminance.
Protect the monitor from dust, for example through appropriate construction measures at
the installation site.
During transport, use the original packaging or service packaging.
Maintain the permissible ambient temperature
The ambient temperature must be in a range of +5 °C to +40 °C.
Avoid reflections on the screen
The monitor has an anti-glare surface which is only effective if the screen is clean and
grease-free.
Comply with the specifications for cleaning.
Position the monitor to avoid reflections on the display area.
Reflections can be caused by lights, windows, furniture with shiny surfaces, or light-col-
ored walls.
In order to reduce reflections on the monitor, only use non-dazzling reflector bulbs for the
ceiling lighting.
Avoid shocks and impacts
The monitor is sensitive to shock. Shocks and impacts on the panel surface can lead to total
failure.
Ensure that such mechanical influences are avoided.
Movable installation
If the monitor is installed such that it can move, make sure that persons or objects in the fa-
cility are not endangered by the monitor's range of movement.