13.1 Information on electromagnetic compatibility (EMC)
Information and manufacturer's declaration – electromagnetic radiation
The CuratOR LX490W is intended for use in the electromagnetic environments noted below.
The customer or user of the CuratOR LX490W has to ensure that the device is used in such an envi-
ronment.
Interference im-
IEC 60601
munity test
Measurement
level
Electrostatic dis-
±8 kV contact
charge (ESD)
±15 kV air
IEC/EN
61000-4-2
Fast transient
±2 kV (5 kHz and
electric distur-
100 kHz burst fre-
bances (burst)
quency)
for power lines
IEC/EN
61000-4-4
±1 kV (5 kHz and
100 kHz burst fre-
quency)
for input/output
lines
Surge voltage
±1 kV line(s)
against line(s)
IEC/EN
61000-4-5
±2 kV line(s)
against ground
Voltage dips,
<5 % V
brief interrup-
in V
tions, and volt-
40 % V
age fluctuations
in V
along power
70 % V
supply lines
in V
IEC/EN
<5% V
61000-4-11
in V
cles (5 seconds)
Magnetic fields
30 A/m
with energy
technology fre-
quencies (50/60
Hz)
IEC/EN
61000-4-8
Note: V
is the alternating current voltage before application of the measurement level.
T
The measurement levels were established according to IEC60601-1-2 4th Edition - Professional
Healthcare.
56
Compliance level
±9 kV contact
±15 kV air
±3 kV (5 kHz and
100 kHz burst fre-
quency)
for power lines
±2 kV (5 kHz and
100 kHz burst fre-
quency)
for input/output lines
±2 kV line(s)
against line(s)
±3 kV line(s)
against ground
(100 % dip
<5 % V
T
) for 0.5 cycles
in V
) for 0.5 cycles
T
T
(60 % dip
40 % V
T
) for 10 cycles
V
) for 10 cycles
T
T
(30 % dip
70 % V
T
) for 25 cycles
V
) for 25 cycles
T
T
(100 % dip
<5% V
T
) for 250 cy-
in V
) for 250 cycles
T
T
(5 seconds)
30 A/m
Information regarding the electro-
magnetic environment
It is recommended to use the device
on wood, concrete, or ceramic floors.
If the floor is made of synthetic mate-
rial, the relative humidity should be at
least 30%.
The power supply quality has to corre-
spond to that of typical industrial envi-
ronments or hospitals.
The power supply quality has to corre-
spond to that of typical industrial envi-
ronments or hospitals.
(100 % dip
The power supply quality has to corre-
T
spond to that of typical industrial envi-
ronments or hospitals.
(60 % dip in
T
If the monitor has to continue opera-
tion even if the power supply is inter-
(30 % dip in
T
rupted, it is recommended to connect
the device to an uninterruptible power
(100 % dip
supply or battery.
T
The magnetic fields with energy tech-
nology frequencies must be in an area
that is representative of a typical loca-
tion in a typical industrial environment
or hospitals.
CuratOR LX490W