Test Procedure 2; Equipment; Test Setup - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Test Procedure 2

Note

Equipment

Test Setup

Agilent E4418B/E4419B Service Guide
This test procedure makes use of a 11683A Range Calibrator fitted
with option H01. The measurement uncertainty of this test
procedure is significantly lower than the previously described
method (refer to Appendix B for a description of the measurement
uncertainty analysis). As such, this test procedure should be used
to ensure the power meter meets it's published specification.
Range Calibrator: Agilent 11683A Option H01
Power Sensor Cable: Agilent 11730A
DC Range Calibrator: Datron 4000A
Power Meter
CHANNEL A
Figure 2-3: Instrument Accuracy (Average Path) Test Setup
Range
Power sensor cable
Calibrator
Performance Tests
Instrument Accuracy Test
DC Calibrator
2-11