0 to 99% settled readings over the GP-IB.
For Agilent 8480 series power sensors
Manual filter, 10 dB decreasing power step:
Number of
Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Normal Mode
Maximum dBm
Maximum dBm
Maximum dBm
150 ms
Typical
Settling
Settling
Settling
200 ms
Times
Times
Times
500 ms
6.6 s
Minimum dBm
Minimum dBm
Minimum dBm
Agilent E4418B/E4419B Service Guide
Table 1-4: Settling Time
1
2
4
8
0.15
0.2
0.3
0.5
0.15
0.18
0.22
0.35
20 dB
20 dB
20 dB
Power
Power
Power
Sensor
Sensor
Sensor
10 dB
10 dB
10 dB
Dynamic
Dynamic
Dynamic
Range
Range
Range
10 dB
10 dB
10 dB
10 dB
10 dB
10 dB
Power Meter Supplemental Characteristics
16
32
64
128
1.1
1.9
3.4
6.6
0.55
1.1
1.9
3.5
x2 Mode
Maximum dBm
150 ms
Typical
Settling
Settling
Settling
180 ms
Times
Times
Times
350 ms
3.5 s
Minimum dBm
Specifications
256
512
1024
13
27
57
6.9
14.5
33
20 dB
20 dB
20 dB
Power
Power
Power
Sensor
Sensor
Sensor
10 dB
10 dB
10 dB
Dynamic
Dynamic
Dynamic
Range
Range
Range
10 dB
10 dB
10 dB
10 dB
10 dB
10 dB