Display Brightness and Contrast Adjustment

Introduction

The following procedure should be performed whenever a front panel
assembly or processor assembly are replaced.
The brightness is controlled automatically after executing the
Set Brightness
Note

Procedure

Agilent E4418B/E4419B Service Guide
softkey, located under the
The contrast adjustment is subjective and varies according to
individual user requirements.
1. Access the
Display
increase the contrast, or the
of the display.
2. When the desired display contrast level is obtained, press the
softkey, located under the
Set Contrast
menu.
Display Brightness and Contrast Adjustment
Service
softkey menu and use the
softkey to decrease the contrast
Service
Adjustments
softkey menu.
softkey to
softkey
3-11