Replaceable Parts
Assembly and Disassembly Guidelines
Replacing the Front Panel Power Reference Cable
Assembly
1.
Remove the front panel from the power meter. (Refer to "Removing the A3
Front Panel Assembly", on page 5-15 for details on removing the front
panel.)
2.
Replace the power reference output cable assembly using the replacement
kit and the torques indicated on the following diagrams.
9 lb/in
5-18
6 lb/in
Agilent E4418B/E4419B Service Guide
Fold up tab
after assembly
25 lb/in