Test Setup; Procedure - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Performance Tests
Instrument Accuracy Test

Test Setup

Procedure

The following procedure should be performed for the Instrument Accuracy
Test.
Note
2-8
Power Meter
CHANNEL A
Figure 2-2: Instrument Accuracy Test Setup
The procedure details the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both channels.
1. Connect the equipment as shown in Figure 2-2.
2. Switch the power meter on.
Preset
3. Press
then
Local
dBm/W
4. Press
,
W
for a reading in watts.
5. Set the range calibrator as follows:
RANGE............................................ 3 µW
POLARITY ...................................... NORMAL
FUNCTION..................................... STANDBY
LINE ............................................... ON
Range
Calibrator
Agilent 11730A
.
Confirm
Agilent E4418B/E4419B Service Guide
POWER
METER

Advertisement

Table of Contents
loading

Table of Contents