Performance Test Record - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Performance Tests

Performance Test Record

Performance Test Record
Model Agilent E4418B/E4419B Power Meter
Tested by ___________________________________________
Serial Number _______________________________________
Date _______________
Test
Zero Test
Instrument Accuracy
3.1623 µW
10 µW
31.6228 µW
100 µW
316.227 µW
1 mW
3.1623 mW
10 mW
31.6228 mW
100 mW
Power Reference
P meas
Interface Test
Functional Test
Low Range, 30 dBm pad
Low Range, 30 dBm pad
Low Range, no pad
Upper Range, no pad
2-30
Table 2-8: Performance Test Record
Min Result
76.40 nW
3.0977 µW
___________
9.904 µW
___________
31.428 µW
___________
99.49 µW
___________
314.72 µW
___________
___________
0.9954 mW
3.1477 mW
___________
9.954 mW
___________
31.477 mW
___________
99.54 mW
___________
0.991 mW
___________
Pass/Fail
___________
Pass/Fail
___________
Pass/Fail
___________
Pass/Fail
___________
Pass/Fail
___________
Channel A
Channel B
Actual
Actual
Result
Result
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
Agilent E4418B/E4419B Service Guide
Max Result
76.40 nW
3.2269 µW
10.096 µW
31.819 µW
100.51 µW
317.74 µW
1.0047 mW
3.1769 mW
10.046 mW
31.769 mW
100.46 mW
1.009 mW
Pass/Fail