Adjustments
Power Reference Oscillator Level Adjustment
14.Remove the power meter's cover and adjust A2R90 until the DVM
indicates the calculated value of V
of A2R90:
Figure 3-4: A2R90 Adjustment Location
Measurement Uncertainty
Refer to Appendix C "Measurement Uncertainty Analysis – Power Reference Level
Test"
3-10
. Refer to Figure 3-4 for the position
1
A2R90
BT1
A2 Assembly
J12
A2J13
Front Panel
Agilent E4418B/E4419B Service Guide
A2L9