Performance Test and Function Test
Performance Test (for Option 007 Temperature Characteristic Test Kit)
Performance Test (for Option 007 Temperature
Characteristic Test Kit)
Introduction
This section provides the information on the performance test setups and procedures for the
E4991A Option 007 Temperature Characteristic Test Kit. To execute the performance tests
for the Option 007, both the Temperature Characteristic Test Kit and the E4991A are
required.
Required Test
The E4991A with Option 007 requires the following performance tests with the test setups
for Option 007 to be done in addition to the performance tests for standard E4991A.
•
Power Level Accuracy Test
•
Impedance Measurement Accuracy Test
Specification
Except for the Power Level Accuracy and Impedance Measurement Accuracy, the E4991A
Option 007 specifications are same as the standard E4991A specifications. The Power
Level Accuracy and Impedance Measurement Accuracy of the Option 007 E4991A are
described below:
Power Level Accuracy:
Impedance Measurement Accuracy
Required Equipment
The required equipment for the Option 007 performance tests is same as that for the
standard E4991A.
Test Procedure
Step 1. Perform all the necessary performance tests for the standard E4991A.
Step 2. Install the E4991A and the hardware of the Option 007 Temperature Characteristic Test Kit
as shown in Figure 2-13. The installation procedure is as follows:
a. Attach the test head of the E4991A to the Test Head Stand (included in the kit).
46
2 dB @ 1 GHz (without Opt. 007)
+2 dB/-4 dB @ 1 GHz (with Opt. 007)
3 dB @ > 1 GHz (without Opt. 007)
+3 dB/-6 dB @ > 1 GHz (with Opt. 007)
The values of accuracy equation terms Ea, Zs and
Yo for Option 007 are different from those of the
standard E4991A. Refer to Specifications on Chapter
11 of the E4991A Operation Manual.
Chapter 2