Instrument Accuracy Test

Specification

Accuracy
1. This performance limit does not include the corresponding sensor power
linearity specification.

Description

The power meter accuracy is verified for various power inputs. There are
two different methods described here that can be used to perform this test,
although one of these methods has a relatively high measurement
uncertainty values associated with it.

Test Procedure 1

Note

Equipment

Agilent E4418B/E4419B Service Guide
Electrical Characteristics
The measurement uncertainty of this test procedure is relatively
high at power levels of 30uW and below, as it uses a standard
11683A Range Calibrator. As such, this test procedure should be
used for indication only.
Range Calibrator ................................... Agilent 11683A
Power Sensor Cable............................... Agilent 11730A
Performance Tests
Instrument Accuracy Test
Performance Limits
1
±0.5% or ±0.02 dB
2-7