1 mW Power Reference

• Power Output:
• Accuracy:
• Frequency:
2
• SWR
:
• Connector Type:
1. National metrology institutes of member states of the Metre Convention,
such as the National Institute of Standards and Technology in the USA, are
signatories to the Comité International des Poids et Mesures Mutual
Recognition Arrangement. Further information is available from the Bureau
International des Poids et Mesures, at http://www.bipm.fr/
2. This SWR specification is only warranted for the E4418B and the E4419B
Power Meters with serial prefix GB4331xxxx and above. Prior to this prefix
the values shown are supplemental characteristics.
Agilent E4418B/E4419B Service Guide
Power Meter Specifications
1
1.00 mW (0.0 dBm).
(for two years)
±0.9% (0 to 55 ºC).
±0.6% (25±10 ºC).
±0.5% (23±3 ºC).
50 MHz nominal.
1.06 maximum (1.08 maximum for
option 003).
Type N (f), 50 ohms.