Removing The A5 Daughter Or A6 Measurement Assemblies5-12 - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A5 Daughter or A6 Measurement Assemblies
Note
1. Connector closed
4. Position sensor flex
circuit into connector
in direction of arrow
5-12
1.
Disconnect the flex circuit from measurement assembly.
Care should be taken when disconnecting the flex circuit from the
measurement assembly.
The flex circuit assembly is released by pushing the connector tab forward
and lifting.
To replace the flex circuit, loop it as shown on page 5-17, and connect the
flex circuit as shown in the figures below.
2.
Slide out the measurement assembly from the side of the power meter.
3.
The daughter assembly is removed vertically.
A5
2. Slide connector clamp
in direction of arrow
5.
6. Push clamp closed in
direction of arrow
Agilent E4418B/E4419B Service Guide
A6
3. Rotate connector
clamp in direction of
arrow