Section3
Setting the Logging Conditions

Sampling Settings

Set the sampling interval, stop conditions, save format to Memory Card, etc.
FUN Mode-[LOGING]-[SAMPLE]
Setting
MODE
1SHOT
REPEAT
STOP (stop conditions)
COUNT
NONE
TIME
FILE
NEW
ADD
OVER WR
CYCLE
ZS-DSU
3-26
User's Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Sets the save method to the Memory Card.
Sampling data is first stored in RAM on the ZS-DSU at short sampling inter-
vals, and is then batch-saved to Memory Card. Data can be logged at a faster
speed at a sampling interval shorter than the REPEAT setting. 1SHOT comes
in handy when testing whether or not measurement has been performed cor-
rectly.
Sampling Timing Chart p.3-27
Sampling data is stored to RAM and saved to Memory Card simultaneously.
This type of logging is not affected by internal RAM size, and is suited to data
logging over an extended period. (default value)
Sampling Timing Chart p.3-28
Sets the sampling stop (start trigger standby state) conditions.
Sets how many times the start trigger to end trigger sequence is to be exe-
cuted. Sampling stops when the preset start trigger to end trigger sequence
count is repeated. (default value: 1)
In this state, sampling (start trigger standby state) is performed at all times.
Samples is not stopped.
Stops sampling when the specified time elapses.
Set how the file (logging data) is to be written to Memory Card.
A new file is created at each start trigger. (default value)
A new file is created at the 1st start trigger after the state changes from a
stopped state to start of sampling, and the data is added to the same file from
the 2nd start to end trigger sequence onwards. Data is added to the same file
until the state changes to a stopped state.
A new file is created at the 1st start trigger after the state changes from a
stopped state to start of sampling, and the same file is overwritten with the
data at the 2nd start trigger. The same file is overwritten with data until the
state changes to a stopped state. When the file is overwritten, the original log-
ging data is erased.
Sets the sampling interval in data logging.
RANGE: When 1SHOT is selected, 150µs to 1 hour
When REPEAT is selected, 10 ms to 1 hour