Omron ZS-DSU41 User Manual page 67

Smart sensors data storage unit
Hide thumbs Also See for ZS-DSU41:
Table of Contents

Advertisement

■ Timing Charts
ONE SHOT mode (high-speed logging)
Sampling
interval
ON
BUSY output
OFF
• The next start trigger is not accepted until saving of the file to Memory Card is completed. When
the save is completed, BUSY output turns ON.
• Data can be sampled at high speed during logging as data is buffered to internal RAM on the ZS-
DSU without the Memory Card being accessed.The size of data that can be saved continuously
without any omissions is affected by the size of the internal RAM on the ZS-DSU.
The following tables show an approximate guide to sampling intervals and logging time:
• When only ZS-LDC only is con-
nected
Num-
ber of
Shortest
Chan
Sampling Interval
nels
1
150 µs
2
200 µs
3
300 µs
4
350 µs
5
400 µs
6
450 µs
7
500 µs
8
550 µs
9
650 µs
10
700 µs
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
End
Start
Logging
Stored in RAM
Batch-saved to
on the ZS-DSU
Memory Card
Sampling data
Data to be saved in a file
Num-
ber of
Chan
Longest
nels
Logging Time
1
10 mins
2
6.5 mins
3
6.5 mins
4
5.5 mins
5
5 mins
6
5 mins
7
4.5 mins
8
4.5 mins
9
4.5 mins
10
4.5 mins
(typical examples)
Setting the Logging Conditions
Time
Start
Logging start trigger
End
Logging end trigger
When connected to ZS-MDC
Shortest
Longest
Sampling Interval
Logging Time
350 µs
400 µs
450 µs
500 µs
550 µs
600 µs
650 µs
650 µs
700 µs
750 µs
(typical examples)
Section3
20 mins
12 mins
9 mins
8 mins
7 mins
6.5 mins
6 mins
5 mins
5 mins
4.5 mins
ZS-DSU
3-27
User's Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zs-dsu

Table of Contents