Omron ZS-DSU41 User Manual page 63

Smart sensors data storage unit
Hide thumbs Also See for ZS-DSU41:
Table of Contents

Advertisement

■ DATA WINDOW
This item is for setting the change in the state of the data (measured value) as the log-
ging end trigger. Set the upper/lower limit value. When the measured value falls inside
or outside of that range, this is taken to be the end trigger.
FUN Mode-[LOGING]-[TRIGGER]-[END]-[WINDOW]
Setting
SOURCE
UPPER
LOWER
MODE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setting the Logging Conditions
Select the source to be used as the logging end trigger.
Range: SourceA to J
Sets the upper limit value of the specified range.
Range: -999.999 to 999.999
Sets the lower limit value of the specified range.
Range: -999.999 to 999.999
Select either of the measured value inside or outside the specified range
is to be used as the end trigger.
Range: IN, OUT (default: IN)
At IN
Logging end
At OUT
Logging end
Section3
Description
Measured value
Specified range
User's Manual
ZS-DSU
3-23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zs-dsu

Table of Contents