AC POWER SOURCE

6.11 Instrument Commands

6.11.1 FETCH and MEASURE Subsystem

Differences between FETCH and MEASURE are: FETCH reads existing measurements
value. Execution time is shorter because data is available immediately but the data may
not be the newest. Instrument will send computer the next sampling data after receiving
MEASURE command. There is time delay so 100ms between MEASure command and
query command is recommended. MEASure command is suitable for power supply output
status changed and wants to get updated measurement value.
Following are tree diagram for FETCH and MEASURE commands:
FETCh | MEASure
:CURRent
:AC?
:1?
:2?
:AMPLitude:MAXimum?
:1?
:2?
:CREStfactor?
:1?
:2?
:INRush?
:1?
:2?
:FREQuency?
:1?
:2?
:POWer
:AC
[:REAL]?
:1?
:2?
:APParent?
:1?
:2?
:PFACtor?
:1?
:2?
:REACtive?
:1?
:2?
79
│ www.deltaww.com
Query the total AC rms current
Query the phase 1 AC rms current
Query the phase 2 AC rms current
Query the total peak current
Query the Phase 1 peak current
Query the Phase 2 peak current
Query the total current crest factor
Query the phase 1 current crest factor
Query the phase 2 current crest factor
Query the total inrush current
Query the phase 1 inrush current
Query the phase 2 inrush current
Query average frequency
Query phase 1 frequency
Query phase 2 frequency
Query total real power
Query phase 1 real power
Query phase 2 real power
Query total apparent power
Query phase 1 apparent power
Query phase 2 apparent power
Query average power factor
Query phase 1 power factor
Query phase 2 power factor
Query total reactive power
Query phase 1 reactive power
Query phase 2 reactive power