5 - 34
Fig. 4
Fig. 5

Camera rotation

WARNING
NOTE
NOTE
ARCOSKOP
System Manual
X-ray radiation!
See Chapter 1, Safety Information.
Use radiation protection! Wear a lead apron!
The position of the VIDEOMED DC to the camera rotation mecha-
nism must be set correct.
See Chapter 6 "Replacing the VIDEOMED DC, Camera rotation 0°
position".
Before adjusting the camera rotation, the keys for horizontal and
vertical image reversal must be off. Camera rotation should be set
to the 0° position. Fig. 4, Fig. 5 and Fig. 6 shows the 0° position.
SPR2-250.840.01
Rev. 01
Adjustments / Programming
Left side of the patient
Right side of the patient
Fig. 6
Seite 34 von 42