5 - 2

mAs counter adjustment

NOTE
Cover the radiation exit window with a lead cover or a folded lead apron.
WARNING
Select the DIAGNOSTICS - MONITORING menu.
Select "MAS COUNTER ADJUSTMENT" in the "Group 1" combo box.
Click on "Update Start".
After about 5 seconds, press the radiation release button on the hand control or the
ARCOSKOP fluoroscopy foot switch.
NOTE
Click on "Update Stop".
Concluding tasks
If no other adjustments or settings need to be made, save the learned data to a disk
(Data menu - Backup on the Service PC).
Perform the checking of the generator adjustment, see next page.
ARCOSKOP
System Manual
The mAs counter adjustment must also be performed after the
generator adjustment.
X-ray radiation!
See Chapter 1, Safety Information.
Use radiation protection.
Wear a lead apron when making adjustments! Maintain a safe
distance!
The collimator closes automatically.
When the mAs-counter adjustment begins, continue pressing the
radiation release button until the message "Stop monitoring"
appears in the "ACTIONS" line. If the radiation release button is
released too soon, the mAs-counter adjustment will have to be
repeated.
A maximum of 9 exposures is automatically released.
When the mAs-counter adjustment has been successfully completed, the mes-
sage "Stop monitoring" appears in the status line.
Adjustments / Programming
SPR2-250.840.01
Rev. 01
09.05
Seite 2 von 42
CS PS 24
5
Siemens AG
Medical Solutions