Loading The Operating Software; Backing Up The System Parameters - Siemens ARCOSKOP Troubleshooting Manual

Table of Contents

Advertisement

Operating software

Loading the operating software

Connect the service PC to the serial service interface on ARCOSKOP.
Insert the "Parameter" disk in drive A.
Start the service program.
Select the Download menu, select the relevant file (Host, Gen....) and start Download.
After successfully downloading, wait for the system to boot (1 minute).
The ARCOSKOP is operational.
NOTE

Backing up the system parameters

NOTE
Prerequisite:
The service PC is already connected, the service program is already started.
Select the "Data" menu.
Select the "Backup" submenu.
Click the "Parameter" check box.
If required select the back-up of further data by selecting the corresponding check box.
Click the "Backup" button. Wait until the data are backed up temporarily on disk.
Then back to the main menu.
Select the "System" menu.
Select the "Logoff" submenu. Wait until the data are stored on disk.
You can now exit the service program by selecting the "System" and "Quit" menus.
Siemens AG
Medical Solutions
If you cannot run Download (e.g. power failure during Download):
- Switch the ARCOSKOP off and back on
- Restart Download
After changing settings / parameters, back these up on the para-
meter disk.
Note that the data are stored correctly only after ending the ser-
vice program
Therefore always exit the service program with "Logoff" and
"Quit".
SPR2-250.840.01
Rev. 01
Seite 1 von 4
09.05
CS PS 24
4 - 1
4
4
4
ARCOSKOP
System Manual