Operating software
Loading the operating software
•
Connect the service PC to the serial service interface on ARCOSKOP.
•
Insert the "Parameter" disk in drive A.
•
Start the service program.
•
Select the Download menu, select the relevant file (Host, Gen....) and start Download.
•
After successfully downloading, wait for the system to boot (1 minute).
•
The ARCOSKOP is operational.
NOTE
Backing up the system parameters
NOTE
Prerequisite:
The service PC is already connected, the service program is already started.
•
Select the "Data" menu.
•
Select the "Backup" submenu.
•
Click the "Parameter" check box.
•
If required select the back-up of further data by selecting the corresponding check box.
•
Click the "Backup" button. Wait until the data are backed up temporarily on disk.
•
Then back to the main menu.
•
Select the "System" menu.
•
Select the "Logoff" submenu. Wait until the data are stored on disk.
•
You can now exit the service program by selecting the "System" and "Quit" menus.
Siemens AG
Medical Solutions
If you cannot run Download (e.g. power failure during Download):
- Switch the ARCOSKOP off and back on
- Restart Download
After changing settings / parameters, back these up on the para-
meter disk.
Note that the data are stored correctly only after ending the ser-
vice program
Therefore always exit the service program with "Logoff" and
"Quit".
SPR2-250.840.01
Rev. 01
Seite 1 von 4
09.05
CS PS 24
4 - 1
4
4
4
ARCOSKOP
System Manual