Table of Contents

Advertisement

5 - 40
POWERPHOS
Centering cross
RBV / I.I.
Fig. 9

Evaluation

Calculate the deviation between the measured area dose product FDPg and the displayed
area dose product FDPa:
The deviation must be < 0.3.
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg)] /
(measured area dose product FDPg).
For deviations > 0.3, the area dose product measuring device must be replaced.
ARCOSKOP
System Manual
Adjustments / Programming
Measuring chamber
Fig. 10
Fig. 11
SPR2-250.840.01
Rev. 01
09.05
POWERPHOS
RBV / I.I.
Seite 40 von 42
CS PS 24