3 - 2

Low voltages

Check the operating voltages of the ARCOSKOP according to tables 1 and 2 (Fig.1 - 4).
From test point
D1.X37 (0V)
D1.X39 (0V)
D1.X39 (0V)
D30.X1.8 (0V)
D30.X21.1 (0V)
D30.X20.1 (0V)
D42.MP1 (0V)
D40.D17.A (0V)
M13.S- (0V)
Tab. 1
*1
With existing "operation free" option, i.e. the control console is detachable, you must mea-
sure and check on board D40 that the voltage does not exceed 5.2 V. Should the voltage
on D40 be too high, the voltage on D42 can be reduced to 5.0 V.

Image intensifier voltages

Take the voltages E1 / E2 / E3 and A from the test protocol of the image intensifier and
check or adjust them at the test points of the control unit listed in Table 2.
Voltage
E1
E2
E3
30 kV, anode
Tab. 2
ARCOSKOP
System Manual
Checking the operating voltages
To test point
D1.X30 (+5 V)
D1.X32 (+15 V)
D1.X33 (-15 V)
D30.X1.9 (+24 V)
D30.X21.10 (+24 V)
D30.X20.3 (+27 V)
D42.MP2 (+5 V)
TP (+5 V)
M13.S+ (+13 V)
Test point
Ground
point (0V)
UE1
UE2
UI 15
UI 30
SPR2-250.840.01
Rev. 01
Voltage
+5.1 V to 5.2 V
+4.9 V to 15.1 V
- 4.8 V to -15.2 V
+24.0 V to 29.6 V
+22.8 V to +29.5 V
+27.3 V to +27.8 V
+ 5,0 V bis +5,2 V
+5,0 V to +5,2 V
+13.1 V to 13.3 V
Potentiome-
Potentiome-
ter for
full format
zoom format
P10
P6
P2
P1
Seite 2 von 4
09.05
CS PS 24
Adjuster
M14. +5 V/Adj.
M14. +15 V/Adj.
M14. -15 V/Adj.
not adjustable
not adjustable
D30 R115
D40.P1
D40.P1 *1
M13.TR1
Voltage divi-
ter for
der ratios
P11
1:1
P7
1:1
P3
1:10000
P1
1:10000
Siemens AG
Medical Solutions
3
3