Reading and storing the load counter data

NOTE
The procedure is divided in 4 steps:
Store the actual load counter values on disk.
Release the resetting of the load counters.
Reset the load counter values to 0.
Return the defective POWERPHOS, including the load counter data disk.

Storing the actual load counter values

Connect the service PC to the serial interface of the unit.
Start the service program.
Select the "Diagnostics / Monitoring" menu.
Select "LOAD COUNTERS" in the "Group 1:" combo box.
Click on "Update Start" and wait until the actual values are displayed.
Click on "Update Stop".
Click on the "Save" button.
Change the data disk of the service program in drive A against a formatted empty one.
Select the drive A for storage.
For the file name, enter the current date, the part no. and the serial no. of the
POWERPHOS.
Example:
The current date is:
Use the date format DD-MM-YY:
The POWERPHOS part no. is:
The POWERPHOS serial no. is:
Enter the file name:
Store the file on disk.
After storage, change the disk in drive A against the previously used data disk of the
service program.
Close the Monitoring window by clicking on the "Cancel" button.
ARCOSKOP
System Manual
Before the replacement of the POWERPHOS, all load counter
values have to be stored on disk. After storing the load counter
values, the load counter has to be reset. All values are set to 0.
After the replacement and all adjustments, send back the disk
with the defective POWERPHOS for repair.
SPR2-250.840.01
Rev. 01
Operating software
08-22-1999
22-08-99
55 73 501
01234
22-08-99_5573501_01234.his
Seite 2 von 4
09.05
CS PS 24
4
4
Siemens AG
Medical Solutions