Prerequisites
Fig. 13
Fig. 15
I.I. service position
Moving the I.I. into the service position
•
Take the threaded rods, spacers and nuts out of the accessory pack ("Service Pack I.I./
Camera" part no. 71 43 519 / Fig. 15).
•
Move the C-arm into the vertical position, with the I.I. below and the POWERPHOS
above.
•
Remove both attachment screws for the I.I. and replace them with the threaded rods
from the service pack.
•
Insert the threaded rods from the service pack at least 5 turns into the threaded holes of
the I.I.
•
Then mount the spacers and nuts (see Fig. 14).
•
Tighten the nuts to fasten the I.I. again.
Siemens AG
Medical Solutions
Fig. 14
SPR2-250.840.01
Rev. 01
09.05