Prerequisites
Fig. 13
Fig. 15

I.I. service position

Moving the I.I. into the service position

Take the threaded rods, spacers and nuts out of the accessory pack ("Service Pack I.I./
Camera" part no. 71 43 519 / Fig. 15).
Move the C-arm into the vertical position, with the I.I. below and the POWERPHOS
above.
Remove both attachment screws for the I.I. and replace them with the threaded rods
from the service pack.
Insert the threaded rods from the service pack at least 5 turns into the threaded holes of
the I.I.
Then mount the spacers and nuts (see Fig. 14).
Tighten the nuts to fasten the I.I. again.
Siemens AG
Medical Solutions
Fig. 14
SPR2-250.840.01
Rev. 01
09.05