Brakes
Fig. 3

Replacement of the angulation brake

CAUTION

Angulation brake magnet

Remove the cover of the horizontal carriage (see Chapter 1 "Covers").
Remove the back cover with the handle for the horizontal lift movement after removing
the 4 countersunk screws. The angulation brake magnet is now visible (Fig. 3).
Remove the 3 attachment screws and remove the brake magnet (arrow/Fig. 3)
NOTE
Remove the plug connection for the brake magnet and insert the new brake magnet.
Place the new brake magnet back on the shaft and attach it with the 3 screws.
Reattach the cable with cable ties.
Reinstall the cover. Make sure the protective ground wire is connected properly.
Perform a function test of the brake (see "Brake forces / Brake torques").
A 30 mm wrench (hexagonal) is required for adjusting the brake.
Siemens AG
Medical Solutions
Danger of burns!
See Chapter 1, Safety Information.
The brake magnet gets hot when the angulation brake is released.
Before touching the brake magnet switch it free of voltage and let
it cool down.
Remove both parts of the brake magnet!
SPR2-250.840.01
Rev. 01
Seite 3 von 6
09.05
CS PS 24
7 - 3