9
Errors

Error List

Error List
The GC error code is reported as CLNNN in which:
C = error class (severity)
L = location
NNN = error number or event number.
The Error class can be one of the following values:
There are five locations:
Table 31 Error list
Error code
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Agilent 990 Micro GC User Manual
0=diagnostic info.
1=advisory warning.
2=recoverable error.
3=critical error.
4=fatal error.
0=mainboard.
1=channel 1.
2=channel 2.
3=channel 3.
4=channel 4.
Class
INFO
INFO
RECOVERABLE
CRITICAL
CRITICAL
RECOVERABLE
INFO
RECOVERABLE
CRITICAL
CRITICAL
CRITICAL
WARNING
INFO
RECOVERABLE
CRITICAL
CRITICAL
CRITICAL
RECOVERABLE
Error messages
Init passed
Init error
Pressure too low, chan = <n>
Pressure too high, chan = <n>
Pressure cannot reach its setpoint, chan = <n>
Pressure restored, chan = <n>
Sample line temperature fault, chan = <n>
Sample line heater sensor fault, chan = <n>
Sample line heater is open, chan = <n>
Sample line heater is short, chan = <n>
Sample line heating too slow, chan = <n>
Sample line temperature is restored, chan = <n>
Injector temperature fault, chan = <n>
Injector heater sensor fault, chan = <n>
Injector heater is open, chan = <n>
Injector heater is short, chan = <n>
Injector heating too slow, chan = <n>
129