7
Replacement Procedures
RTS Filter Replacement
RTS Filter Replacement
Tools required
The following tools are required to perform the replacement procedure described following
section. Allow approximately 5 to 10 minutes to complete this process.
•
3/16 open-end wrench
•
Torx T-10
•
Torx T-20
Replacement procedure for Micro GC RTS filter
1 Shut down the GC, and allow it to cool to ambient room temperature.
2 Remove the main back cover of the GC as described in
3 Remove the channel using the procedure in
4 Unscrew the two Torx T-10 screws and remove the DEGC cover.
100
Removing the rear covers
Channel Exchange and Installation
Agilent 990 Micro GC User Manual
on page 95.
on page 85.