Sample Gas Handling
G3588-68863
7 Use a 5/16 inch wrench to screw the sample line nuts to the Genie filter outlet.
8 Connect the 1/16 tubing to the sample inlet manifold by the 1/16 inch Swagelok fitting.
The pressure regulator has a maximum inlet pressure of 3,000 psi. Applying higher pressures
WAR N IN G
may result in serious personal injury and equipment damage.
9 Connect the Sample IN port on the pressure regulator to the sample input line.
10 Start the GC (see
11 Leak test the system to ensure that all connections are leak free.
52
Long Storage Recovery Procedure
on page 41).