Sample Gas Handling - Agilent Technologies 990 User Manual

Micro gas chromatograph
Hide thumbs Also See for 990:
Table of Contents

Advertisement

Sample Gas Handling

4
The Micro GC is built for the analysis of gases and vapors only. You are advised to prepare a
noncondensing gaseous standard sample for routine checkup of the instrument. Sample
pressure should be between 0 and 100 kPa (0 to 15 psi), the temperature between 0 and 110 °C,
and it must be filtered, preferably through a 5-mm filter. Agilent always recommends the use of
the external filter kit (CP736729) between the injector and the sampling device.
For more details, see
Liquids in sample gases must be avoided because they will critically damage the instrument.
C AU T I O N
Agilent 990 Micro GC User Manual
44
47
48
56
Using the External Filter Unit
49
50
on page 47.