System Power On; Facility Leakage Current And Ground Continutiy Test Notificataion - GE Brivo OEC 715 Service Manual

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

3.5.3.

System Power On

When you place a recently unpacked system into use, allow it to stand in its new
environment long enough for any condensed moisture to evaporate. A decrease in
CAUTION
temperature or a humidity increase can cause water vapor in the air to condense,
sometimes on electric circuits. You may damage the equipment if you apply power
before this condensation dries.
1. Take out footswitch from the accessory package. Connect the footswitch to C-Arm.
2. Connect the interconnect cable to C-Arm.
3. Plug the power cord into appropriate AC receptacle. Turn on the circuit breaker on the back of
Workstation, and then press the power on/off key in front of Workstation.
Before power on the system, please check and make sure that the voltage of the system
WARNING
is compatible with the power supply. Or you must change the power supply settings of
the system.
Refer to Section 3.5.1 for details.
4. Verify that the workstation and C-Arm complete software boot and initialization with no errors
reported on their displays.
The auto fluoroscopy mode is selected and the power on indicator lights. The field size is in
Normal mode.
The left monitor of workstation will display GE logo screen. The right monitor will enter into the
User login Screen automatically.
5. Press vertical lift button and verify that vertical lift column is operational.
3.5.4.
Facility leakage Current and Ground Continuity Test
Notification
The medical facility may require to perform current leakage and grounding continuity tests on the
system before it is put into service. In that case, the leakage current and ground continuity tests should
be performed as required.
Refer to section 8.4.5 to do ground continuity tests.
Notify the service personnel to follow local regulation and following waring:
Ground continuity tests shall be conducted before undertaking any leakage current test.
WARNING
Power off the system before and during any tests. Do not interrupt power to the system
WARNING
at any time while the system is booting or shutting down to avoid file system corruption.
Installation