Disposal Information; Accessing Stored Reports And Rfid Data - Guardian R1000 Operation Manual

Table of Contents

Advertisement

Figure 30: Main Menu screen.
Select "Run Report" to view and
print reports
Figure 31: Select "Browse
Test Results" to see a list of
data records
Figure 32: The last 200 data
records are displayed here.
Press the "Enter" button to view
any report details
Figure 33: Test Results detail
screen. Press the "Print" button
to print the test results; press the
"Enter" button to return to the
Data Records screen

DISPOSAL INFORMATION

• Any materials exposed to potential biological threat agents dur-
ing testing should be disposed of as any other suspected BIO-
HAZARD waste. Consult local regulations for proper disposal
procedures.
• Any Guardian Reader which delivers a positive result, later
confirmed by a certified laboratory, will be replaced at Alex-
eter's expense.
• All materials supplied by Alexeter (test strips, buffer solution
and sampling materials) are non-toxic and may be disposed of
in the ordinary waste stream. Consult local regulations.
DANGER: The Guardian Reader System™ is used to identify poten-
tial biohazards. Death or serious injury may result from exposure
to biohazardous materials. Handle all suspect material with proper
safety equipment.
ACCESSING STORED REPORTS & R.F.I.D.
DATA
Up to 200 previous test reports are stored in the Reader's non-
volatile memory. These test report data files can be recalled from mem-
ory, displayed on the LCD screen and re-printed. After a total of 200
data files have been stored on the Guardian Reader, data from new test
files overwrite previous data files. The oldest data file is overwritten
with the newest data file so that the most recent 200 data files always
reside in the Reader's memory. Loss of power due to non-volatile
battery drainage or dislodging of the battery itself may result in loss
of stored data. Regular backups of stored data are recommended (See
"Reader Data Export", pg. 27).
To view and print stored reports follow these steps:
1)
From the MAIN MENU, select "RUN REPORT" using the
UP (∆) or DOWN (∇) arrow keys (Figure 30). Press the
"ENTER" button.
2)
Select "BROWSE TEST RESULTS" using the arrow keys
(Figure 31). Press the "ENTER" button.
3)
The display will list (up to) the last 200 tests you have run.
The time, date and sample number are listed for each previ-
ous test on the "DATA RECORDS" screen, beginning with
the most recent test (Figure 32). Using the arrow keys, select
the desired Test Report. Press the "ENTER" button to view
the test details.
4)
The selected Test Report will be displayed on the LCD
Page 19