Centering the X-ray tube / Image intensifier
Final steps
•
Remove jumpers at plug X18 .
•
Reinstall the I.I. collision protection.
•
Reinstall the shock wave head cover.
•
Measure the protective conductor function.
•
Perform an IQ test according to chapter 6 "Dose rate" of the Start-up Instructions (does
not apply when replacing the shock wave head).
•
Replace or supplement the documents.
•
Check all functions.