RUSKA 7750i
Users Manual
Changing the Number of Decimals
Each unit has a default number of decimal places used for pressure display. This may be adjusted
up or down by one decimal place.
1. The decimal digits are set from the Menu | Menu |Setup | User. From the Main Menu (press
PREVIOUS until the Main Menu appears), press Menu [F6], Menu [F6], then Setup [F2], and
then User [F2]. Press the MODE key to toggle between the Ps and the Qc channels.
2. Turn the rotary knob until the label "Display digits" is highlighted.
3. Turn the rotary knob further until the desired number of digits from the default resolution is
highlighted (-1, 0, +1), and then press the ENTER key to select.
Key Click
The ADTS can be configured to emit a "clicking" sound each time a key is pressed.
1. The key "click" is set from the Menu | Menu |Setup | User. From the Main Menu (press
PREVIOUS until the Main Menu appears), press Menu [F6], Menu [F6], then Setup [F2], and
then User [F2].
2. Turn the rotary knob to highlight "Key click" and then highlight either on or off.
3. Press the ENTER key to select.
Menu | Setup — Units
In addition to the standard units of measure provided by the ADTS, two user-defined units are
available. To create a user-defined unit, the user enters a name that is one to ten characters long
and a conversion factor that is a multiple of kiloPascals (kPa).
For example, using the information from Table 2-1, the conversion factor for millitorr or one
micron of mercury at 0 °C is calculated as follows:
mTorr
Thus the conversion factor simplifies to
4-14
1000
mTorr
=
kPa
x
1
Torr
Figure 4-12. Menu | Menu | Setup | Units
1
Torr
1
mm
Hg
x
x
. 0
0193377
oC
1
mmHg
0
mTorr =
kPa
x
7500
o
0
. 0
1450377
psi
x
psi
1
kPa
.
6180
gme19.bmp