Qc Channel - Fluke RUSKA 7750i User Manual

Air data test set
Table of Contents

Advertisement

RUSKA 7750i
Users Manual
The screen will display the current status of the pressure sensor Stability and
Temperature.
If any of the above are unstable, then the system will delay until stability is achieved.
Pressing OK [F6] will bypass this wait period.
Bypassing this wait period can have a negative effect on the zeroing
procedure.

Qc Channel

1.
2.
3.
4.
5. Wait for the zeroing procedure to finish.
6.
6-8
Verify that the Ps and Qc sensors are close to atmospheric pressure.
Enter the Calibration screen by selecting MENU | MENU | CALIBRATION.
Select Zero [F1]. Do NOT press the Calibrate button.
Figure 6-3. Menu | Menu | Calibration
Do not disturb the instrument while zeroing is in process.
Press PREVIOUS to return to the Main Menu.
Note
Figure 6-4. Zeroing Screen