Fluke RUSKA 7750i User Manual page 100

Air data test set
Table of Contents

Advertisement

RUSKA 7750i
Users Manual
Specifications
General Specifications
Pressure Range:
Internal Volume:
Custom Ranges:
Display:
Display Resolution:
Electrical Power:
Operating Temperature:
Storage Temperature:
Humidity:
Dimensions:
Weight:
Standard Pressure Units:
Pneumatic Ports:
Relief Valves:
Supply Pressure:
Supply Vacuum:
Ingress Protection:
EMC:
Electrical Safety:
Pressure Safety:
Interface:
A-6
Standard available ranges;
Ps: 32 inHg Abs. (0 to 1085 mbar Abs.) or,
40 inHg Abs. (0 to 1355 mbar Abs.)
Qc: 32 inHg Diff. (0 to 1085 mbar Diff.) or,
68 inHg Diff. (0 to 2300 mbar Diff.)
Ps: 24 ± 1 in³
Qc: 1.8 ± 0.3 in³
Consult Factory
6.4 inch, TFT Active Matrix Color
1 ft; 0.1 m; 0.1 km/h; 0.1 knots; 0.00001 Mach
100 – 120 VAC, 220 – 240, VAC 50/50 Hz, 150 W max.
18 °C to 36 °C
–20 °C to 70 °C
5% to 95% relative humidity, non-condensing
Nominal Control Volume: 5 to 60 in
7 in H x 16.5 in W x 19 in D (17.8 cm x 41.9 cm x 48.3 cm)
45 lb (20 kg)
Engineering Units: inHg at 0 °C and 60 °F, kPa, bar, psi,
inH
O at 4 °C, and 25 °C, kg/cm2, mmHg at 0 °C,
2
cmHg at 0 °C, and cmH
units of measure.
Aeronautical Units: feet, meters, knots, km/h, Mach
1/4 inch NPT female, with AN4 & AN6 adapters supplied.
Test Port: 120% of Maximum Scaled Quartz Sensor Range
60 psi ± 5 psi (4 bar ± 0.5 bar)
Minimum 50 liters per minute with auto vent feature,
Ultimate vacuum less than 1 mtorr.
IP-20, Indoor use, Altitude <2000 in
EN 61326
EN 61010
Pressure Equipment Directive 97/23/EC-class,
Sound Engineering Practice (SEP)
IEEE 488 and RS232
3
(80 to 1000 cm³)
O at 4 °C & two user defined linear
2

Advertisement