The test access password allows the user to protect access to ADTS configuration and programs.
If the test access password is set to any number other than zero (factory default), it is required
before the user is allowed to change the limits, control parameters or programs. The access key is
only shown on the Ps channel. Press MODE key to toggle between the Ps and the Qc channels.
It is recommended that the access password be recorded and filed in a secure
location.
Menu | Setup — User
The Menu | Menu | Setup | User menu is used to setup all of the user specific configurations.
This includes setting the control step size, range of bar-graph, pressure gas head, display pressure
filter, display resolution digits, and audible key click. The step size, bar-graph max, ready
tolerance and gas head are defined separately for each channel. All other entries are common
for both channels. Press MODE key to toggle between the Ps and the Qc channels.
Step Size
In addition to entering the pressure set point by the key-pad, the pressure set point can be changed
using the step function. The step function is used primarily when the operator is taking pressure
steps in equal pressure increments. The size of the pressure step is user defined. From the Main
Menu (press PREVIOUS until the Main Menu appears), press Menu [F6], Menu [F6], Setup [F2],
and then User [F2]. Using the rotary knob, move the cursor to highlight Step Size. Enter the
desired step size in the current pressure units and press ENTER.
Bar Graph Maximum
The bar graph on the Main Menu screen can be scaled to match the device under test by setting
the full scale value of the bar graph.