Test 7: Variable Baseline (Option 001)
Note
Test Specifications
Equipment Needed
Procedure
This t,est is only to be performed if Option 001 is installed.
Range:
Accuracy:
1. Digitizing Voltmeter (DVM)
2. High Power attenuator
3. 508, O.l%, 1OW Feedthrough
1. On the HP 8114A set up the OUTPUT page as shown in the
following illustration:
2. Connect HP 8114A to the DVM as shown:
Equipment Set-up 1 for Variable Baseline Test
edings for the following baseline settings (VB). Multiply the r
3. Set the DVM HP 3458A to:
Function: DCV
Trigger: TRIG EXT
AD-Converter integration time NPLC: 0.1
(NPLC = Number of Power Line Cycles)
Configuring the Output Page