Test 1: Period; Test Specifications; Equipment Needed; Procedure - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

Test 1: Period

Test Specifications

Equipment Needed

Procedure

Range
Resolution
Accuracy
RMS-Jitter
Counter
Cable, 50 62, coaxial, RNC
1. On the HP 8114A press TRIGGER and set up page as follows:
, '-' .I,j.rjgj,e F;ul.ses
2. On the HP 8114A press OUTPUT and set up page as shown in the
following illustration:
3. Connect the HP 8114A to the Counter as follows:
HP 81 i 4A UNDER TEST
Connecting HP 8114A to the Counter
4. Set the Counter to:
FUNCTION
INPUT A
66.7
ns to 999 ms
3 digits, best case 100 ps
0.03% + 25 ps (0.05% + 25ps for period ~100 ns)
Configuring the Output Page
Period A
50 0
Testing the HP 8114A