Duty-cycle

Standard HP 8114A (Baseline = 0 V)

0.1% to 100%
(Subject to width and period specifications).
1 0
Amplitude (50 Ohm into 50 Ohm)
Figure 2-1. Duty-cycle / Amplitude Ranges
Figure 2-l shows the maximum possible duty-cycle for a given pulse
amplitude from 5052 into 5061. Note that amplitude doubles from HIZ
(High-Z) into 508.
In double-pulse mode the actual duty-cycle of the signal is twice the
value displayed on the HP 8114A screen because two pulses are
generated per pulse period. Therefore, the duty-cycle available, and
set, will be limited to half the
20
3 0
40
value
given by Figure 2-1.