Amplitude

Warning

Baseline:

Variable Baseline (Option 001)

Range:

1.00 V to 50.0 V (doubles into open circuit) 2.00 V to 100 V (HIZ
(High-Z) into 5OQ)
the output voltage can be dangerous to life. Care should
therefore be taken when connecting the HP 8114A to external
instruments.
When working in HIZ (High-Z) Mode, if you remove
the external load the output voltage can be higher than the
programmed voltage. V,, can be as much as 130 V, even when
set as low as 2 V,,.

Current:

40.0

Accuracy:

Resolution:

3 digits, best case 10 mV
0 V ilO mV 50.5% of amplitude
Range:
-25.0 V to +25 V
Accuracy: