Procedure - HP 8114A Service Manual

Programmable pulse generator 100 v/2 a
Table of Contents

Advertisement

A) Amplitude Test (from 509

Procedure

2. On the HP 8114A set up the UUTPUT page as shown in the
following illustration:
3. Set HP 8114A Delay to
4. 'kke the baseline reading as B,,a
5. Set HP 8114A Delay to 50 ms and Amplitude to 1 V
6. 'E&e the high-level reading as H,,Q
7. Calculate the Amplitude:
A
8. Repeat steps 3 to 7 for the following settings:
into 503)
Configuring the Output Rage
0
ns
Amplitude limit
Amplitude
minimum
maximum
0.89 v
1.11 v
2 v
1.38 v
2.12 v
5.15 v
4.86 v
10 v
9.80 v
10.2 v
20.3 V
20 v
19.7 v
50.6 V
49.4 v
7
TR Entrj
6 - l
6-2
6 - 3
6-4
6 - 5
6 - 6
Testing the HP 8114A